+
  • 1(1748243964834).jpg

Biosafety cabinet


1. The vertical laminar flow negative-pressure system achieves 70% air recirculation plus 30% directional exhaust, equipped with a zero-leakage HEPA/ULPA dual-effect filter to maintain ISO Class 4-5 cleanliness. 2. The double-layer SUS304 stainless steel negative-pressure enclosure integrates a fully sealed 4L liquid collection tank; the inner liner, formed in one piece with rounded corners, features both leak-proof and self-cleaning properties.
Product Categories: Biosafety cabinet
  • Product Description
  • Structural Design:
    1. The vertical laminar flow negative-pressure system achieves 70% air recirculation plus 30% directional exhaust, equipped with zero-leakage HEPA/ULPA dual-effect filtration to maintain ISO Class 4-5 cleanliness.
    2. The double-layer SUS304 stainless steel negative-pressure housing integrates a 4L fully sealed liquid collection tank. The inner liner, with its seamlessly formed rounded corners, combines leak-proof performance with self-cleaning properties.
    3. The 10°-inclined operating surface, combined with a detachable hand rest, forms an ergonomic system, complemented by a maintenance-free, imported, height-adjustable glass front window (≥5mm).

     

    Intelligent Control:
    1. The intelligent airflow compensation system ensures that when the filter resistance increases by 50%, the airflow fluctuation remains below 10%.
    2. The Chinese-language human-machine interface provides real-time display of parameters such as wind speed, filter life, and operating time, and incorporates a dual-warning mechanism for both clogging and malfunctions.
    3. Triple safety interlock: Front window limit switch with audible and visual alarm linked to lighting, UV sterilization interlock, and filter differential pressure protection.
     

    Security Expansion:
    1. The unobstructed return-air structure optimizes airflow organization, while the splash-proof power supply and multi-valve interface meet diverse experimental needs.
    2. The imported fan works in conjunction with a multi-sealing system to achieve a stable intake air velocity of 0.2 m/s ±15%.

     

    Model HKHJ-1100IlA2-XHKHJ-1383llA2-XHKHJ-1500IIA2-X
    Cleanliness Level HEPA; ISO Class 5 (100 Class 100) / ULPA: ISO Class 4 (10 Class 10)
    Filter Category/Filter Efficiency HEPA: ≥99.995%, @0.3μm / ULPA: ≥99.999%, @0.12μm
    Downwind speed (m/s) 0.35
    Inflow wind speed (m/s) 0.55
    Noise (dB(A)) ≤65
    Power supply AC220V/50HZ/1Ph
    Average airflow
    Biological protection
    Personal Protective Equipment a. The total number of colonies collected by the impact chamber sampler ≤ 10 CFU per sample. b. The total number of colonies collected by the slit-type sampler ≤ 5 CFU per sample.
    Protective performance of the test product Total colony count ≤ 5 CFU per test.
    Cross-infection prevention Total colony count ≤ 2 CFU per test.
    Maximum rated power (including backup outlets) 2.0 kVA 2.1 kVA 2.1 kVA
    Weight (kg) 260300350
    Workspace dimensions (mm) 940*600*660 1210*600*660 1350*600*660
    External dimensions (mm) 1100*750*2250 1383×775×2295 1500*750*2250
    Fluorescent lamp specifications (W) and quantity 36 times 2 36 times 2 36 times 3
    UV lamp specifications (W) and quantity 20*1 20*1 40*1
    Illuminance (Lx) ≥800≥800≥800
    Exhaust air direction Eject

Previous

None

Next

None

Related Products

Product Information


Submit